| Moving distance | 600 mm |
| Moving speed | 100 mm/s |
| Selectable speeds | 0.2 - 100 mm/s |
| Positioning accuracy | ± 0.25 mm |
| Projection accuracy | ± 0.5 mm at 4 m / 13 ft |
| Line length | 3 m at 10 ft, 6 m at 20 ft (red); 2.7 m at 9 ft (green) |
| Power supply | 110 - 230 VAC |
| PC Interface | RS485 |
| Dimensions | floor mount rail: 1695 × 143 × 117 mm ceiling mount rail: 1812 × 143 × 117 mm |
| Feedback control | absolute linear encoder |
| Calibration | auto calibrating |
| On-board data processing | microprocessor |
| Hardware requirements for ISOMARKnova | PC with WINDOWS XP |
| Simulation data download | 1-5 networks, sequentially |
Subject to modifications |
|
| LASER MODULES. |
 |
|
| | DORADO red | DORADO green |
| Line width (up to 4 m distance) | < 1 mm (focusable) | < 1 mm (focusable) |
| Line length (at 3 m / 10 ft distance) | 6 m / 20 ft | 2.7 m / 9 ft |
| Laser type | Diode | Diode Pumped Solid State |
| Wavelength | 635 nm | 532 nm |
| Beam Color | red | green |
| Output power | < 1 mW | < 1 mW |
| Laser class | 2 | 2 |
| Drift | not measureable | not measureable |
| Ambient temperature | 0 - 40 C | 15 - 30 C |
subject to modifications |
|
|
|
|
DORADOnova systems are built failsafe.
The built-in circuit and linear encoder checks for possible stepper motor misses which, without independent verification, could go undetected.
The control system compares the computed isocenter location with the laser position on the rail.
The laser will only switch on after the control system has verified that the lasers have reached the prescribed rail positions.
These automatic control procedures assure that the isocenter location is accurately marked at all times. |
 |
|
|
|